51
Signametrics
iLevel
int
Level indicator. 0 – low, 1 – high.
Return Value
The return value is one of the following constants.
Value
Meaning
SCAN_OKAY
Operation successfully completed.
Negative Value
Error code
Example
status = SCANSetTriggerOut(0,1); //Set trigger line to high level
SCANSetupStep
H/W access Command
;
Polled Command
Description
Set the Scanner for Stepped scanning operation.
#include “SM4040.H”
int
SCANSetupStep(int
iScan
)
Remarks
This function prepares the Scanner for a single step scanning operation. See
SCANStep
function for details. It clears the hardware ScanList pointer to point to the first entry of
the ScanList table. Step() command increments this pointer.
Parameter
Type/Description
iScan
int
Identifies the Scanner being addressed.
Return Value
Integer error code.
Value
Meaning
SCAN_OKAY
Operation successfully completed.
Negative Value
Error code
Example
int status = SCANSetupStep(0); // Set for step scan operation
SCANStep
H/W access Command
;
Polled Command
Description
Step to the next channel in the Scan List.
#include “SM4040.H”
int SCANStep(int
iScan
)
Remarks
Switch to the next point in the scan table. This is software triggered switching function. It
is similar to
SCANTriggerScan
except the scanner steps to the next channel in the scan
list due to a command rather than by hardware event. The number of points is the actual
number of times
SCANStep
is issued following
SCANSetupStep
command.
SCANSetupStep
sets the hardware ScanList pointer to point to the first entry of the
ScanList table. On
SCANStep
command, the Scanner selects the channel(s) pointed to
by the hardware pointer, then increments the ScanList pointer. Parameters effecting the
operation include the set configuration, the Actuation time, and trigger output polarity
Summary of Contents for SM4020