SC5412A Operating & Programming Manual
Rev 1.1.0
4
•
Ground yourself using a grounding strap or by touching a grounded metal object.
•
Touch the antistatic bag to a grounded metal object before removing the hardware
from its packaging.
•
Never
touch exposed signal pins. Due to the inherent performance degradation
caused by ESD protection circuits in the RF path, the device has minimal ESD
protection against direct injection of ESD into the RF signal pins.
•
When not in use, store all SignalCore products in their original antistatic bags.
G
E T T I N G
S
T A R T E D
Unpacking
All SignalCore products ship in antistatic packaging (bags) to prevent damage from electrostatic
discharge (ESD). Under certain conditions, an ESD event can instantly and permanently damage several
of the components found in SignalCore products. Therefore, to avoid damage when handling any
SignalCore hardware, you must take the following precautions:
Remove the product from its packaging and inspect it for loose components or any signs of damage.
Notify SignalCore immediately if the product appears damaged in any way.
Verifying the Contents of your Shipment
Verify that your SC5412A kit contains the following items:
Quantity Item
1
SC5412A IQ Modulator
1
USB Flash Drive Installation Software (may be combined with other products onto a single drive)
1
Getting Started Guide
Setting Up and Configuring the SC5412A
The SC5412A is a designed for use in a PXIe or PXIe hybrid chassis. Chassis manufacturers must provide
at least the minimum required per-slot power dissipation cooling capability to be compliant with the
PXIe specifications. The SC5412A is designed to be sufficiently cooled in either all-PXIe chassis or PXIe
hybrid chassis (PXI Express chassis with traditional PXI slots). However, certain environmental factors
may degrade performance. Inadequate cooling can cause the temperature inside the RF housing to rise
above the maximum for this product, leading to improper performance and potentially reducing product
lifespan or causing complete product failure. Maintain adequate air space around the chassis at all
times, and keep the chassis fan filters clean and unobstructed.
Refer to your chassis manufacturer’s user manual for proper setup and maintenance of your
PXIe or PXIe hybrid chassis. The SC5412A on-board temperature sensor should indicate a rise
of no more than 20 °C above ambient temperature under normal operating conditions.
!
!