background image

7.To open the slide clamp, hold the tubing 

down at the keyhole while pulling the slide 
clamp up and out as in Fig. 2.  Open the 
roller clamp and confirm no flow.

When the pump is turned on and programmed 
infusion data exists in memory, a screen  
displays asking the operator if the intended 
use for the pump is for a New Patient.

Ÿ

Answering 

YES

 to this prompt clears the 

existing infusion data.

Ÿ

Answering 

NO

 retains the data and allows 

the operator to resume the infusion.

Using the soft arrow keys, scroll to the desired 

care area and press 

OK

.

1.Type the drug’s first two letters (similar to 

phone texting).  A list of all matching drugs 
will automatically be shown. 

2.Using the soft arrow keys, scroll to the 

desired drug and press 

OK

.

IMPORTANT: 

 If a drug is not available in the 

Master Drug Library, follow your facility’s 
procedure to request a drug library update.

New Patient Prompt

Select Your Care Area

Select a Drug

For drugs NOT in drug library

Basic Infusions

3

2

(Set Loading Continued)

(Basic Infusion Continued)

1.Select  the Care Area.
2.Enter “B” “A” and scroll to the BASIC selection.
3.Using the soft arrow keys select Primary or Secondary Bag and 

press OK.

4.Select a dose mode and press 

OK 

(default is mL/hr).

5.Type in the required values, press 

OK 

after each entry.

6.Press 

RUN/STOP 

to start the infusion.

7. Check flow to confirm clamps are open, no tubing kinks and drops 

are falling.

1.Program a Primary Infusion.  
2.Lower the Primary Bag at least 20” below the Secondary Bag.
3.Stop the pump if running, press 

REVIEW/PROGRAM 

soft key.

4.Press 

PROGRAM SECONDARY

 soft key.

5.Type the drug’s first two letters.
6.Press 

OK 

to select/confirm secondary bag.

7.Follow all setup steps.
8.Press 

RUN/STOP 

to begin Secondary infusion.

9.Check flow for drops falling in the secondary not the primary drip 

chamber and press 

YES/NO 

soft key to confirm.

1.Close the roller clamp on the set below the pump
2.To unload set, insert the slide clamp in keyhole at       to open the 

door and clamp the tubing.

3.Remove the tubing by pulling out from the bottom towards the top.

Secondary Infusions

Set Unloading

1

2

3

LOAD SET

4

Figure 1

Figure 2