43
SIGMA Manual Number 41018
Revision AR
PUMP OPERATION
Operating Tips
Follow all prompts on the pump screen.
■
Always press
OK
to confirm entries and changes.
■
Follow all prompts on screen to clear alarms.
Be sure to load sets properly.
■
Start with a fresh section of tubing loaded in the pump, make sure there is enough tubing for
patient movement.
■
To open the pump door, insert the gravity IV set’s slide clamp fully into the keyhole.
■
Load tubing tautly, from top to bottom in loading points 1, 2, 3 and 4, following the red/
green prompts.
■
Close the door with thumb pressure over both door hooks, which is at the top of the door
and the bottom of the door.
■
Open the slide clamp by pulling it straight up, while holding the tubing around it down.
Use the Drug Error Prevention (DEP) system.
■
DEP system protects against human errors that could cause Adverse Drug Events.
■
BASIC mode can not detect human errors.
Do not drop the power supply.
■
The power supply is an electronic device. It is not simply a plug, and it will break if repeat-
edly dropped.
■
Keep the power cord plugged in whenever the pump is not in use.
Follow secondary procedures.
■
Use drop hooks to drop primary containers below secondary containers.
■
With secondary rates above 300 mL/hr, look for and clamp off primary line siphoning.