![Sigma 3-30KS Operating Manual Download Page 30](http://html1.mh-extra.com/html/sigma/3-30ks/3-30ks_operating-manual_1267903030.webp)
Refrigerated Centrifuge Sigma 3-30KS
5 Set-up and connection
30 / 93
Version 07/2012, Rev. 1.22 of 02/06/2021
• sb
Translation of the original operating manual, part no. 0701302
5.2.2
Customer-provided fuses
Typically, centrifuges with a nominal voltage of 220-240 V must be
protected with 16 Amp B fuses that are to be provided by the customer.
Centrifuges with a nominal voltage of 200 V or 208 V must be protected on
site with a 16 Amp type D circuit breaker.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.