![Sigma 3-16K Operating Manual Download Page 63](http://html1.mh-extra.com/html/sigma/3-16k/3-16k_operating-manual_1267896063.webp)
Operating Manual SIGMA 3-16K, page 63 of 69
03/06
7.2.4
Emergency Lid Release
At the upper right panel there is a stopper which can be removed eg. with a
screw driver. The lid can be released by pulling the visible string.
Attention!
The lid may only be unlocked and opened when the rotor is at standstill.
7.2.5
Problems with the Centrifuge?
Please contact your supplier for support or in the event of malfunctions and for
supply of spare parts.
7.3
Error Mode
In the error mode the Start-, Stop- and Lid-keys are flashing. “ERROR“ is
displayed in the speed area and the error code in the rcf area.
7.3.1
Error Codes
Error no.
Kind of error
Actions
1
Tacho signal disturbed
•
Power off/on
4
Lid does not open after
pressing the Lid-key
•
Power off/on
•
Emergency lid release
2, 3, 5 – 11, 17 - 30 Internal fault
•
Power off/on
12
Imbalance determined
•
Remove imbalance
13
Temperature sensor
•
Contact service
14
Excess temperature (>50°C)
•
Let centrifuge cool down
Should it not be possible to repair the failure, please contact Service!
7.4
Speed-RCF-Diagram
An additional help is the enclosed Speed-RCF-Diagram.
Summary of Contents for 3-16K
Page 1: ... Refrigerated Centrifuge 3 16K Operating Manual ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 10: ......
Page 28: ...Operating Manual SIGMA 3 16K page 28 of 69 03 06 ...
Page 38: ...Operating Manual SIGMA 3 16K page 38 of 69 03 06 ...
Page 66: ......
Page 68: ......
Page 69: ...Operating Manual SIGMA 3 16K page 69 of 69 03 06 ...