7 Malfunctions and error correction
Version 07/2015, Rev. 1.12 of 16/07/2019
• sb
37 / 68
Translation of the original operating manual, part no. 0702802
7
Malfunctions and error correction
7.1
General malfunctions
Malfunctions are indicated by error messages with a number in the
speed/RCF display. In addition, the blue LED flashes rapidly.
In the event of a fatal error (e.g. a defective lid lock), a certain safety time
will be counted down on the display. During this time, ―ERR‖ and ―SAFE‖
flash alternately on the display. When the time is up, "OFF" will be
displayed.
WARNING
Do not switch the centrifuge off until ―OFF‖ is displayed! This is necessary
in order to ensure that the rotor is at a complete standstill.
• Eliminate the source of the problem (see tables below).
• Acknowledge the error messages with the lid key.
Type of error
Possible reason
Correction
No indication on the display
No power in the mains supply
Check fuse in the mains supply
Power cord is not plugged in
Plug in power cord correctly
Mains power switch off
Switch mains power switch on
Lid is not closed correctly
Close the lid
Centrifuge cannot be started: The set
speed value is displayed in an
unchanged manner
Several possible causes
Power off/on. If the error occurs
again, contact service
The lid lock is not closed correctly
Open and close lid. If the error
occurs again, contact service
Centrifuge decelerates during
operation and displays an error from 1
to 18 after powering on
Several possible causes
Power off/on. If the error occurs
again, contact service
Centrifuge decelerates during
operation and displays error 19 after
powering on
Several possible causes
Acknowledge by pressing the lid key
Lid cannot be opened
Lid locks have not released
Unlock the lid manually (see chapter
7.1.1 - "Emergency lid release")and
contact service
Lid seal sticks
Clean the lid seal and apply talcum
powder
Temperature value cannot be reached
(only for refrigerated centrifuges)
Condenser dirty
Contact service