Laboratory Centrifuge Sigma 2-16P
5 Set-up and connection
Version 04/2007, Rev. 2.23 of 28/04/2022
• sb
29 / 76
Translation from the original operating manual, part no. 0700602
Pos : 93 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050- 0020-0020 Sic herungen baus eits ( außer 3-30KS, 8KS, 8KBS) @ 7\m od_1306312302397_68.docx @ 43513 @ 3 @ 1
→
5.2.2
Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp B fuses that are to
be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.