Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
10 Technical data
Version 11/2017, Rev. 1.7.1 of 08/12/2022
• sb
65 / 113
Translation of the original operating manual,
→
10 Technical data
Manufacturer
Sigma Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Type:
4-5KRL
4-5KRL IVD
Part no.:
91309, 91564
100008, 100009
Connection requirements
Electrical connection:
Protection class:
IP code:
see name plate
I
20
Power consumption (kW):
Input fuse (AT):
Series fuse at the transformer (AT):
1.5
12.0 (temperature fuse)
16.0 (temperature fuse, 1-phase)
Performance data
Max. speed (rpm):
Max. capacity (ml):
Max. gravitational field (x g):
Max. kinetic energy (Nm):
4,700
840
4,470
20,078
Other parameters
Time range:
Temperature range:
Storage locations:
10 sec
– 99 h 59 min 59 sec,
short run, continuous run
-10°C to +40°C
50
Physical data
Height (mm):
Height with open lid (mm):
Width (mm):
Depth (mm):
Weight (kg):
406
890
677
680
135
Noise level (dB(A)):
< 59 (at maximum speed)
Refrigerant data
Refrigerant:
Global warming potential (GWP):
Filling quantity (kg):
Max. permissible pressure (bar):
CO
2
equivalent (t):
R134a
2,140
0.500
23
0.650