Refrigerated Centrifuge Sigma 1-16K
Refrigerated Centrifuge Sigma 1-16K IVD
8 Maintenance and service
44 / 70
Version 10/2013, Rev. 2.21 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0700302
→
8.1.2.1
Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.4 - "Resistance data").
• If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
WARNING
Plastic accessories must not be greased!
8.1.3
Rotors
The rotors are produced with the highest precision, in order to withstand the
permanent high stress from high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
• Check the material regularly (at least once a month) for
– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.
• Check the bores of the rotors and multiple carriers.
• Replace any damaged components immediately for your own safety.
8.1.4
Microhaematocrit rotor
• Remove the microhaematrocrit rotor for cleaning.
• Wipe the centrifuge chamber clean.
• Replace the rubber ring in the event of wear or glass breakage (part no.
16003 for rotor 11024).