SIGLENT
SSG3000X Service Manual 27
Test Procedures
1. Synchronize the SSG3000X and signal analyzer. Connect the RF output
terminal of the SSG3000X with the RF input terminal of the signal analyzer.
Connect the USB DEVICE connector of the SSG3000X with the USB
interface of the PC as shown in figure.
2. Install the EasyIQ software to the PC. Configure the IQ modulation wave
table using the EasyIQ softwave.
3. Configure the wave Data Source
、
the modulation type of the wave
、
the filter
parameter of the wave.
Then download the currently generated wave to the SSG3000X.
4. Configure the SSG3000X:
a) Set the frequency to 800 MHz.
b) Set the level to 3 dBm.
c) Press I/Q to enter the I/Q source settings interface.
d) Enable the I/Q state switch.
e) Set the source to internal.