SSA5000A Quick Guide 9
Cooling Requirements
This instrument relies on the forced air cooling with internal fans and ventilation
openings. Care must be taken to avoid restricting the airflow around the
apertures (fan holes) at each side of the analyzer. To ensure adequate
ventilation it is required to leave a 15 cm (6 inch) minimum gap around the
sides of the instrument.
CAUTION:
Do not block the ventilation holes located on both sides
of the analyzer.
CAUTION:
Do not allow any foreign matter to enter the analyzer
through the ventilation holes, etc.
Power and Grounding Requirements
The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC
power at 50/60 Hz (+/-5%), or single-phase 100 to 120 Vrms (+/-10%) AC
power at 400 Hz (+/-5%).
No manual voltage selection is required because the instrument automatically
adapts to line voltage.
Depending on the type and number of options and accessories (probes, PC
port plug-in, etc.), the instrument can consume up to 193 W of power.
Note:
The instrument automatically adapts to the AC line input within the
Summary of Contents for SSA5000A