SDS5000XSeries Digital Oscilloscope Service Manual
46/ 90
WWW.SIGLENT.COM
3.
Press the Measure button on the front panel of the oscilloscope to
display Peak-Peak measurement.
4.
Set waveform of the 9500B to sine, amplitude to 30mV, and
impedance to 50 Ω
, and set the impedance of the oscilloscope
channels to
50 Ω
.
5.
Set frequency of the 9500B to the frequency in
Table 7
Frequency points in the test
Scope Model
Impedance
Frequency Point (Hz)
SDS5104X
50
Ω
0.05M,21M,52M,74M,98M,121M,168M,201M
,251M,301M,351M,401M,451M,501M,601M,
701M,801M, 901M, 1001M, 1002M, 1201M,
1401M, 1601M, 1801M, 2001M
SDS5054X
50
Ω
0.05M,21M,52M,74M,121M,168M,201M,234
M,268M,301M,334M,367M,401M,434M,468
M,500M,601M,801M
SDS5034X
50
Ω
0.05M,21M,52M,74M,121M,168M,201M,234
M, 268M,301M,316M, 331M,351M,371M,
451M
SDS5104X
1 M
Ω
0.05M,21M,52M,74M,98M,121M,168M,201M
,251M,301M,351M,401M,451M,484M,501M,
601M
SDS5054X
1 M
Ω
0.05M,21M,52M,74M,98M,121M,168M,201M
,251M,301M,351M,401M,451M,484M,501M,
601M
SDS5034X
1M
Ω
0.05M,21M,52M,74M,98M, 168M, 201M,
251M, 301M, 351M, 401M, 501M