SDS5000XSeries Digital Oscilloscope Service Manual
37/ 90
6.
Make a linear regression with these 5 data points:
Offseterror1~Offseterror5, Offsetout1~Offsetout5. Check to verify that
the slope is within the limits specified in
7.
Check to verify that abs(Vzero) in step 4 is within the limits specified
8.
Set the oscilloscope Volts/Div to the other settings in
, and
repeat the steps 4 to 6.
9.
Turn the 9500B output off.
10.
Disconnect the connections to all the channels.
Table 4
Offset Accuracy
Volt/Div
Offset
Slope
error
Zero
input(V)
10 V/div
35 V,30 V,0,-3 0V,-35 V
±1.5%
±1.202
5 V/div
30 V,25 V,0,-25 V,-30 V
±1.5%
±0.602
2 V/div
20 V,10 V,0,-10 V,-20 V
±1.5%
±0.242
1 V/div
10 V,5 V,0,-5 V,-10 V
±1.5%
±0.122
500 mV/div
5 V,2.5 V,0,-2.5 V,-5 V
±1.5%
±0.062
200 mV/div
2 V,1 V,0,-1 V,-2 V
±1.5%
±0.026
100 mV/div
1 V,0.5 V,0,-0.5 V,-1 V
±1.5%
±0.014
50 mV/div
500 mV,250 mV,0,-250 mV,-500
mV
±1.5%
±0.008
20 mV/div
200 mV,100 mV,0,-100 mV,-200
mV
±1.5%
±0.0044