SDS5000XSeries Digital Oscilloscope Service Manual
66/ 90
WWW.SIGLENT.COM
7
Solving General Problems
This chapterincludes suggestions for solving general problems.
If there is no display
Is the light of the power button glowing?
If yes:
Go to the Troubleshootingchapter and check the power supply
modules and acquisition board.
If no:
Check that the power cord is secured in the power receptacle.
Check that the fuse in the power receptacle is not blown out.
Check that the power source is alive.
If there is still no light glowing, go to the Troubleshootingchapter
and check the power supply modules and acquisition board.
If the display is stuck at the logo screen
Remove all the USB FLASH drivers plugged into the oscilloscope
and power cycle the oscilloscope to try again.