Copyright ©Sigicom AB 2003-2015
13
17.
If filter weight is 3.0 mg or more, change filter. See section
4.2.2 for a detailed description.
18.
If you changed the filter:
i) press RESET.
ii) press ENTER.
iii) press START three times (password).
iv) press ENTER to confirm filter change.
If not:
i) press RESET twice.
19.
Connect your flow meter to the Osiris inlet located above
the display.
20.
Hold the flow meter vertical. Adjust the flow by pressing
or
until flow meter shows 600 ml/min.
21.
When flow rate is correct, press ENTER, and record the
change in your log.
22.
Remove flow meter and press RESET.