the D10/D10 Micro status and settings, and make changes of configurable
parameters both for the D10/D10 Micro and its sensors.
When any configuration change is made in INFRA Net, for example the
communication schedule, it will be applied to D10/D10 Micro at its next
communication. Then:
1.
Monitoring is stopped.
2.
parameter(s) is/are changed
3.
monitoring is restarted.
Some parameters (e.g. the INFRA Net communication schedule), are
changed without monitoring is stopped and restarted.
When the instrument is in monitoring off mode, the INFRA Net Remote
system can also be used, however physical access to the D10/D10 Micro
is required. Use the following procedure:
1.
Perform a manual communication (upload) on the D10/D10 Micro to
send the current status and settings to INFRA Net (see section 6.5.2).
2.
Wait for the status to be updated on INFRA Net (Hardware tab).
3.
Use INFRA Net to change one pr several parameters.
4.
Perform a manual communication again on the D10/D10 Micro to
read the new change request. The changes will then be performed.
6.13.1.
Force Reboot
It is possible to force a reboot (= restart) on an D10/D10 Micro which is
monitoring. This is done by selecting (in INFRA Net – Hardware tab)
Special commands → Force Reboot
followed by
Manage changes → Commit.
D10/D10 Micro will perform the reboot at the next communication.