Appendix A Specifications
A.1 Reference conditions (factory calibration)
Range/Standard:
FAST; 55 - 130 dBA / S04
Sound pressure level: 94 dB
Frequency:
1000 Hz
Temperature:
23 ºC
Static-pressure:
101.3 kPa
Direction:
90 degrees off the sound source
A.2 Accuracy
S50 follows the demands for class 1 sound level meters, which
implies that it is capable to measure the sound pressure level of a
sinusoidal signal at the calibration frequency, for each frequency
weighting provided, within an absolute accuracy better than ±0.7 dB
(ANSI) and extended uncertainty (95% confidence) of 1.1 dB (IEC).
Transient file low-pass filter characteristics (except for frequency
weighting):
+/- 0.25dB up to 1250 Hz
-3dB at 1500Hz
-25dB at 2000Hz
Warm-up time: Less than one minute after power-up of the
controlling Datalogger.