PVA-3000 Reference Manual
December 2, 2019
Sifos Technologies
Page
41
3.
PVA Interactive Software
PVA Interactive
is the graphical user interface for the PhyView Analyzer. It is a Tcl/Tk application that is built from
a number of Tcl/Tk scripts. It exclusively uses the PowerShell PSA API for all instrument interactions and control
meaning that anything that can be done in PVA Interactive can also be done in PowerShell PSA.
3.1.
Overview of Menus
PVA Interactive consists of two menus that are always present.
The PVA Interactive
Main Menu
provides full access to all test
port resources including port configuration, link configuration,
impairment control, metering, calibrations, and automated
testing. Test port subsystems are all organized by category on
tabbed sub-menus that allow arbitrary navigation through test
port resources. These subsystems include:
Port & Impairment Configurations
Link Configuration
Link Monitor
PVA Meters
PVA Packet Flow
PHY Performance Tests
PVA Meter Calibrations
Generally, the Main Menu always operates on the currently selected test port within the currently connected PVA
instrument.
Separately, a
Slot-Port Select
menu allows selection of any available test port in the currently connected instrument.
This means that the test port can be altered independent of any tab menu selections in the Main Menu at any time.
3.2.
Initial PVA Connection
When PVA Interactive is opened, the user will be prompted to select an instrument to which PVA Interactive will
initially connect
(see Figure 3.2)
. The default selection will be the
most recently
connected PhyView Analyzer. This
brief dialog assures that various users on a common network sharing
multiple instruments can connect to an instrument without interfering
with other instruments that might be in use by other host computers. It
also assures that any
selected
analyzer to be utilized is powered up and
connected to the local network.
The
PVA Chassis Selection
dialog (
Figure 3.2
) verifies the presence of
any selected or entered PVA address, inventories the selected instrument
for available PVA test ports, and allows users to add any new
instruments that may have been recently connected to the network.
Note
:
If the PVA Chassis Selection dialog
fails to find a connected and
available PVA
instrument
, it will ultimately open PVA Interactive in
Demo Mode
. When the application opens in Demo Mode, the
annotation
*DEMO MODE*
will appear in red on the Slot-Port selection
menu. Demo Mode is described in Section 3.14 below.
3.3.
The PVA Main Menu
After selecting and connecting to a PVA Chassis address, the
Main Menu
(
see Figure 3.3
) will appear along with the
Slot-Port Control helper menu. The Main Menu will open by default to the first tab sub-menu,
Port Config
. Each of
the tab sub-menus will be addressed in Sections 3.6 through 3.13 below.
Figure 3.2
PSA Connection Dialog
Figure 3.1
PVA Interactive 4.2 Startup
Summary of Contents for PhyView PVA-3000
Page 22: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 22 ...
Page 70: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 70 ...
Page 108: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 108 ...
Page 124: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 124 ...