Sifam Tinsley Insu 10 User Manual Download Page 6

0

+23C ± 2 K

Insu 10

12. Specifications

Meas.

Range

Resolu-

Accuracy

Overload

function

tion

(..% of rdg

value and

+

 

.. digit)

duration

± 

15

6

Insulation Resistance Measurement

Note

During Insulation resistance measurement the           symbol 
flashes as long as the high voltage is present at the terminals

1) For insulation resistance range, 2) For low Ohms/continuity Ranges
*

Terminal voltage on open

* Open circuit voltage

circuit (DC)

5V  ± 1V d.c.

- 0% to + 30% of rated voltage * Lead resistance

*

Short circuit current < 2 mA

compensation : 0 - 9.99 

Ω

*

Test current on load 1 mA at
minimum pass values of
insulation as specified
in VDE 0413 Part 1.

Reference Conditions :

Ambient Temp.
Relative Humidity

45%...55%

Battery Voltage

8V ± 0.1V

Voltage Measurement

AC (Sine), 50/60 Hz

Display

LCD display field (65 mm x 30 mm) with analog indication and Digital 
display and with display of unit of measured quantity and functions

Analog

Display

LCD scale with bar graph pointer

Scale Length

47 mm

Scaling

0...30 with 30 graduations

Over range Indication

By triangle (13)

Digital

Display/Char. Height

7 segment digits/12mm

Number of Digits

3 digit for M

Ω

 and 

Ω

4 digit for Stop watch

Over range Indication

OL

ON

2 0

R

Insulation
Resistance

1)

M

Ω

U = 50 V,

N

100 V

0.01 M

Ω

 to

0.99 M

Ω

≥1.0 M

Ω

 to

9.9 M

Ω

≥10 M

Ω

 to

99 M

Ω

10 K

Ω

(

)

0.01 M

Ω

100 K

Ω

 to

(

)

0.1M

Ω

1 M

Ω

± 3% ± 2

± 5% ± 2

± 30%

1200V rms,

10 sec

Insulation
Resistance

1)

M

Ω

U = 250 V,

N

500 V,
1000 V

0.01 M

Ω

 to

9.99 M

Ω

≥10.0 M

Ω

 to

99.9 M

Ω

≥100 M

Ω

 

to

999 M

Ω

99.9M

Ω

10 K

Ω

(

)

0.01 M

Ω

100 K

Ω

 to

(

)

0.1M

Ω

1 M

Ω

± 3% ± 2

± 5% ± 2

± 30%

1200V rms,

10 sec

2)

Low Ohms 

Ω

0-9.99 

Ω

≥10

 Ω

-99.9 

Ω

0.01 

Ω

 at

210 mA

0.01

 Ω

 at

21 mA

± 3% ± 2

± 5% ± 2

1200V rms,

10 sec

2)

Continuity  

0-9.99 

Ω

≥10

Ω

-99.9 

Ω

0.01 

Ω

 at

210 mA

0.01

Ω

 at

21 mA

± 3% ± 2

± 5% ± 2

1200V rms,

10 sec

V AC/DC

1 V

± 2% ± 3

1200V rms,

10 sec

≥25V-450V

≥450V-600V

1 V

± 3%

50V

100V

250V

500V

1KV

V

M

W

Summary of Contents for Insu 10

Page 1: ...am Tinsley Instrumentation Inc 3105 Creekside Village Drive Suite No 801 Kennesaw GA 30144 USA Contact No 1 404 736 4903 E mail Id psk sifamtinsley com Web www sifamtinsley com User Manual Analog Digital Insulation Tester Insu 10 ...

Page 2: ...poses the operator the system and the measuring instrument to danger The use of repaired fuses or short circui ng of the fuse holder is prohibited 13 3 Housing No special maintenance is required for the housing Excessive contamination has an adverse effect on isolation and reduces input resistance The surface must be kept clean for this reason Use a slightly dampened cloth for cleaning Oil free pe...

Page 3: ...SE appears on display If display OL Fuse is OK If display FUSE Fuse is blown and needs replacement Note Once FUSE is displayed all the functions are inhibited Immediately switch the Instrument OFF and replace the fuse 18 3 1 Liquid Crystal Display 10 Symbols for displaying the 2 On Off push button selected functions 3 Push button 11 Display for selected function 4 Push button 12 Display for the un...

Page 4: ...STOP WATCH mode or in MENU structure During an Insulation test connections must not be touched After insulation tests capacitive circuits must be allowed to discharge before disconnecting test leads Test leads and crocodile clips must be in good condition clean and with no broken or cracked insulation u 4 17 Power Supply Battery 6 x 1 5V cells IEC LR6 Nickel cadmium rechargeable cells may be used ...

Page 5: ...the measuring circuit and ensure high voltage is not available on leads Automatic turn off Your instrument turn off automatically if any of the keys or the function selector switch have not been activated for about 10 minutes in insulation resistance range and 5 minutes in other ranges How to prevent automatic TURN OFF In order to prevent automatic TURN OFF select CONTINUOUSLY ON mode For this pus...

Page 6: ...ication and Digital display and with display of unit of measured quantity and functions Analog Display LCD scale with bar graph pointer Scale Length 47 mm Scaling 0 30 with 30 graduations Over range Indication By triangle 13 Digital Display Char Height 7 segment digits 12mm Number of Digits 3 digit for MΩ and Ω 4 digit for Stop watch Over range Indication OL ON 2 0 R Insulation Resistance 1 MΩ U 5...

Page 7: ...ow the last reading for few seconds even after releasing the TEST key Note The instrument s batteries are depleted during insulation resistance measurement Only press and hold the TEST key as long as it is necessary to take the reading in the manual measurement mode Continuous measurement as described below should be performed if absolutely necessary 7 14 11 1 Description of Menu Items 11 1 1 Sett...

Page 8: ... is set it is applicable to all the 5 ranges of Insulation measurement n Select the required value of the time for which the test is to be carried out n Briefly Press the TEST key The insulation resistance test carries on for the time interval selected n The test terminates automatically after the selected time has elapsed n The display then flashes the value of measurement at the end until the Key ...

Page 9: ...u can still read the actual measured value Apply the measured quantity to instrument prior to activating MIN MAX function The MIN MAX function is switched off when 1 Key is pressed for approximately for 1 Sec OR 2 Then function selector switch is operated OR 3 When the instrument is turned OFF and ON again 9 12 8 Voltage Measurement V AC DC You can use this instrument for voltage measurement from ...

Page 10: ...nal if the measured resistance value is less than the pre setted GoNoGO value Default value is 0Ω Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resistance by zero adjustment Connect the test leads to instrument and connect the free ends of the leads ...

Reviews: