QUICK START GUIDE

Redy

Industrial

TM

 Series

1 |  RedyIndustrial_v1_2022

Safety and General Information

1.   For safe operation, please consider the process and ambient conditions for which the device is designed and specified at the time of 

delivery. This information can be found on the data label on the instrument housing and in the Instruction Manual.

2. Do not remove the electronics housing to prevent damage to the unit. A damaged hologram seal will void the warranty.
3. There are no serviceable parts inside the unit. Repairs must be performed by qualified personnel only.

Installation and Electrical Connection

1.   Do not use Teflon® tape or liquid sealing on the process connections. 
2. Please ensure that the piping is clean and purged before you connect the flow meter.
3. The gas used must be clean and dry.
4.  For particle protection, we recommend to install our 50µm filter at the inlet.
5.  It is recommended that the user checks the device on a regular schedule to ensure that it is leak free as both metal and elastomeric seals, 

gaskets, O-rings and valve seats may change with age, exposure to process gas.

 

  

Note

  Your Sierra sales representative can help you to source suitable O-rings, elastomeric seals, fittings and filters.

6.  Please refer to the RedyIndustrial Instruction Manual for detailed instruction on mounting orientation and location. 
7. Please verify a proper electrical connection. The device must be grounded. The supply voltage is 18-30 vdc (typically ±50 mV).

 

  

Note

  The current consumption depends on the instrument version. Please refer to the information in the Instruction Manual.

This Quick Start Guide for the RedyIndustrial Series Thermal Mass 

Flow Meters and Controllers 

gives easy set up and installation 

instructions. For complete instructions, please download and read the 
RedyIndustrial Series Instruction Manual. Before installing meter, follow 
the steps below.

Step 1. Download Instruction Manual

To get technical information, download the RedyIndustrial Series  
Instruction Manual at 

sierrainstruments.com/RedyIndustrial-manual

Step 2. Download FREE Software

Download the most current RedyIndustrial product software at 

sierrainstruments.com/RedyIndustrial-software

. Use the included cable  

(if ordered)  to connect the meter to your computer.

IMPORTANT!

  If you do not have an internet connection, you must 

download the software to a USB stick or other storage device.

Default Factory Settings

Modbus RTU
Communication speed

9600 Baud

Data bits

8

Stop bits

2

Parity

None

Node address

247

Profibus

 (optional)

Node address

125

When using several devices simultaneously the user has to 
configure each device with an individual address!

Profinet / EtherCAT

 (optional)

Name

RedyIndustrial SIP

IP address

0.0.0.0 or 192.168.0.50

Further information on the device configuration can be found in the 
Instruction Manual.