Output Signal Wiring

All flow meters are equipped with either a calibrated 0–5 VDC, 0–10
VDC or 4–20 mA output signal. These linear output signals represent
0 to 100% of the flow meter user full scale.

DC Output Wiring

The 0-5 VDC or 0-10 VDC output signal can drive a minimum load of
1000 Ohms. For 0-5 VDC or 0-10 VDC connections, connect to the
terminals marked Output signal and Signal ground.

4–20 mA Output Wiring

The 4-20 mA current loop output is self-powered (non-isolated).
The maximum loop resistance (load) for the output is 400 Ohms. For
a 4–20 mA connection, connect to the terminals marked Output sig-
nal and Signal ground.

Troubleshooting the Flow Meter

Begin hardware troubleshooting by verifying the following facilities is-
sues are correct. These areas impact system operation and must be
corrected prior to performing any flow meter inspections.

1.  Verify the incoming power to the flow meter is present and of the

correct voltage and polarity.

 
2.  Check the flow meter wiring for correct connections.
 
3.  Verify the flow meter is installed with the correct number of up-

stream and downstream pipe diameters as shown in Figure 2.

 
4.  Verify the flow direction indicator is correctly aligned pointing

downstream of flow.

 
5.  Make sure there are no leaks in the line being measured.
 
After verifying the factors above, follow the troubleshooting procedures
outlined on the next page. If you need to return the flow meter to the
factory, see page 10 for return shipping instructions.

Flow Meter Calibration

Sierra Instruments maintains a fully-equipped calibration laboratory.
All measuring and test equipment used in the calibration of Sierra me-
ters are traceable to NIST standards. Sierra is ISO-9001 registered and
conforms to the requirements of ANSI/NCSL-Z540 and ISO/IEC
Guide 25. If the flow body or electronics have been damaged or you
simply want to have the flow meter re-calibrated, contact the factory
for return shipping instructions. Calibration must be performed by
qualified personnel using NIST-traceable equipment.

Caution!

Do not connect power to

the 4-20 mA terminals.

Caution!

Before attempting any flow

meter repair, verify that the

line is not pressurized.

Caution!

Always remove main power

before disassembling any

part of the mass flow meter.

Summary of Contents for Accu-Mass 730

Page 1: ...Part Number IM 73 AM Revision E 05 04 5 Harris Court Building L Monterey CA 93940 831 373 0200 800 866 0200 Fax 831 373 4402 http www sierrainstruments com Sierra Instruments b v Bolstoen 30A 1046 AV...

Page 2: ...publication may be copied or distributed transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form or by any means electronic mechanical manua...

Page 3: ...bstructed Flow Requirements 5 Installation 6 Wiring Connections NEMA 2 Enclosures 7 Wiring Connections NEMA 4 Enclosures 7 Output Signal Wiring 8 Troubleshooting the Flow Meter 8 Returning Equipment t...

Page 4: ...meter repair verify that the line is de pressurized Caution Always remove main power before disassembling any part of the flow meter Caution Printed circuit boards are sensitive to electrostatic disc...

Page 5: ...of a 304 stainless steel probe which is inserted in the gas stream The sensor consists of two sensing ele ments a velocity sensor and a temperature sensor which automati cally corrects for changes in...

Page 6: ...ial Do not return any equipment to the factory without first contacting Sierra Customer Service Technical Assistance If you encounter a problem with the flow meter review the configura tion informatio...

Page 7: ...measured 5 For remote installations verify the supplied cable length is sufficient to connect the flow meter sensor to the remote electronics Do not extend or shorten the supplied cable be tween the...

Page 8: ...umber of diameters of straight pipe required downstream of the flow meter Flow Flow Flow Flow Flow Flow Figure 2 Recommended Pipe Length Requirements for Installation Installation The following instru...

Page 9: ...input power 300 mA load maximum to pin 1 and power ground to pin 5 See Output Sig nal Wiring for signal wiring instructions Figure 3 NEMA 2 CE Enclosure Wiring Connections Warning All wiring procedure...

Page 10: ...s 3 Verify the flow meter is installed with the correct number of up stream and downstream pipe diameters as shown in Figure 2 4 Verify the flow direction indicator is correctly aligned pointing downs...

Page 11: ...flow Correct alignment with the flow indica tor pointing downstream in the direc tion of flow No response to flow from sensor assembly No power Turn on power to the flow meter Flow profile distortion...

Page 12: ...ts com Or contact Customer Service at service sierrainstruments com Email or 800 866 0200 or 831 373 0200 in the US or 31 0 20 6145810 in Europe Return shipments to USA Headquarters Sierra Instruments...

Page 13: ...DC 300 mA Output Signal Linear 0 5 VDC or 0 10 VDC proportional to mass flow rate 1000 Ohms minimum load resistance or linear 4 20 mA proportional to mass flow rate 400 Ohms maximum resistance Perform...

Page 14: ...ite Signal 3 Red DC input Power 4 N C 5 N C 6 Green Power Ground 7 Brown or Orange Bridge Black Connector Female Mating Connector Solder Side No Color Signal 1 Red DC input Power 2 White Signal 3 Blac...

Reviews: