Uroskop ACCESS
SPL5-330.840.01
Page 10 of 12
Siemens AG
Endoscopy Option
Rev. 01
03.03
CS SD 24
Medical Solutions
5 - 10
Error Correction
Example of the test configuration with a chrome-plated Allen wrench, contrast OK.
Arrow 1:
The reflective areas of the Allen wrench appear almost bright and glaring and have a
noticeable white border, indicating the beginning of camera overmodulation. The reflective
areas still indicate noise in the image.
Arrow 2:
When the monitor is set correctly, the base structures are visible without saturation
effects.
Fig. 8