Siemens
7482255 Rev 01
Page 9 of 12
October 2002
Chapter 1 - Introduction and System Requirements
3.2.1 Temperature Specifications
The G50 and the G60 S can be operated in the environment listed in Table 1-5.
3.2.2 Humidity Specifications
The G50 and the G60 S are designed to operate in an environment that can
experience humidity conditions ranging from 30% to 85% relative humidity,
non-condensing. During storage and transport, the G50 and the G60 S can
withstand from 0% to 95% relative humidity, non-condensing.
3.2.3 Altitude
The G50 and the G60 S design can operate at altitudes from 1060 hPa (-400 m
or -1312 feet) to 700 hPa (3050 m or 10,000 feet).
!
Caution:
The G50 and the G60 S and the installed peripherals (OEMs) contain
cooling fans that are required to maintain the equipment at acceptable operating
temperatures. Over-temperature conditions can impair the accuracy of the sys-
tem and damage its components. Do not attempt to operate the G50, the G60 S,
or the installed peripherals if any fan becomes inoperable.
See "Diagnostic Indicators" (page 6-9): LED 8 illuminates when the temperature
in the E module exceeds the over-temperature limit of 80 °C (176 °F). Consult
the
Instructions for Use
for information on the warning message displayed when
operating temperatures are exceeded.
Table 1-5
Temperature Specifications
Specification Value
System Operating Temperature
(a)
10 °C to 40 °C
(50 °F to 104 °F)
System Transport Temperature
-10 °C to 55 °C
(14 °F to 131 °F)
Array Transducer Operating Temperature
10 °C to 40 °C
(50 °F to 104 °F)
Array Transducer Transport Temperature
-10 °C to 60 °C
(14 °F to 140 °F)
Mechanical Sector Transducer Operating Temperature
20 °C to 40 °C
(68 °F to 104 °F)
Mechanical Sector Transducer Transport Temperature
-5 °C to 50 °C
(23 °F to 122 °F)
Pencil CW Transducer Operating Temperature
10 °C to 40 °C
(50 °F to 104 °F)
Pencil CW Transducer Transport Temperature
-10 °C to 60 °C
(14 °F to 140 °F)
(a)
Operation of the MO drive is guaranteed to 35
°C
(95
°F
)
Summary of Contents for SONOLINE G50
Page 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Page 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Page 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Page 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Page 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Page 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Page 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Page 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Page 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Page 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Page 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...