Print No.:

Siemens Medical Solutions

Ultrasound Division

ejt

SONOLINE G20

System

Installation Instructions

US

2004

US06-101.812.01.02.02

Replaces: US06-101.812.01.01.02

English
Doc. Gen. Date: 10.04

n.a.

Part No.: 

sd

SONOLINE G20 System Delivery Procedure

08647815

08647831

Summary of Contents for Sonoline G20

Page 1: ...ll rights including rights created by patent grant or registration of a utility model or design are reserved Print No Siemens Medical Solutions Ultrasound Division ejt SONOLINE G20 System Installation Instructions US 2004 US06 101 812 01 02 02 Replaces US06 101 812 01 01 02 English Doc Gen Date 10 04 n a Part No sd SONOLINE G20 System Delivery Procedure 08647815 08647831 ...

Page 2: ...icrosystems Inc Adobe Acrobat is a registered trademark of Adobe Systems Inc Other products and brand names are trademarks of their respective owners Disclaimer The service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie mens or one of its affiliates to provide such services Assemble...

Page 3: ...ther equitable relief will be appropriate to restrain such use threatened or actual Buyer further agrees that i any of Siemens s suppliers of software is a direct and intended beneficiary of this end user sublicense and may enforce it directly against Buyer with respect to software supplied by such supplier and ii No supplier of Siemens shall be liable to buyer for any general special direct indi ...

Page 4: ...___________ 5 Uncrate the system and peripherals 6 Prior to Starting 6 Required Materials 6 Unpack the System 6 Verify the Receipt of Accessories 9 Inspect the System 9 Power On the System 10 Front Control Panel 14 Verifications 15 Verify the Monitor Configuration 15 Verify Transducer Port Operation 16 Verify the Imaging Modes 17 Verify Printer Operation 19 Verify VCR Operation if installed 21 ...

Page 5: ...k for the SONOLINE G20 ultrasound system The courtesy check includes Uncrate the System and Peripherals Power On the System Verifications The time required to perform this procedure is approximately one hour NOTE In the event that the uncrating or courtesy check cannot be com pleted or the customer refuses the uncrating or courtesy check immediately contact the Siemens regional sales representativ...

Page 6: ...ltrasound system is not designed for lifting instead transport it on its casters Prior to Starting 0 Allow sufficient space to remove the system from the crate Required Materials 0 Utility knife Diagonal cutters Unpack the System 0 1 Check the shipping container for damage caused during shipment NOTE Notify Siemens Uptime Service Center if damage to the shipping container is found 2 Remove the shi...

Page 7: ...he top of the bars down and away from the clip To remove the clip pull the clip away from the cardboard box Fig 1 Corro Clip Removal Pos A Squeeze together to release lock Pos B Pull away from the shipping container 5 Unpack the shipping crate Remove the accessory box Remove the cardboard box from the G20 system Remove and retain the screws that secure the two roll out ramps to the cardboard base ...

Page 8: ...hat is placed on the floor and two holes at the end that is placed on the ramp stud Fig 3 Ramp Plank Placement 7 Unfold and lay flat the cardboard shipping base Fig 4 Shipping Base Cardboard 8 Remove the plastic cover from the ultrasound system 9 Roll the system off the shipping pallet Disengage the brakes on each wheel Carefully roll the system off the shipping pallet down the ramps previously in...

Page 9: ... Inspect the System 0 1 Verify the brake pedal is functional Press the brake pedal down to engage the brakes Press the brake pedal up to disengage the brakes Fig 5 Brake Pedal Pos A Press the brake pedal up to disengage the brake Pos B Press the brake pedal down to engage the brake 2 Verify the system is free of cosmetic damage including scratches and blemishes 3 Position the ultrasound system in ...

Page 10: ... with the small safety lock part number 4H540092 0DB For 230 V systems use the gray power cord part number HG14A13V3 5M with the large safety lock part number 4H540070 0DB Attach the safety lock to the system power inlet as shown in the figure below Insert the lower post of the safety lock first then the upper post Fig 6 Power cord Safety Lock Pos A Safety lock Pos B Power inlet Pos C Lower post o...

Page 11: ...on the MAINS power switch up is on and down is off The Standby button is backlit with the color amber when the MAINS power is switched to on Fig 7 Back of the System Power Pos A MAINS power switch Pos B Power cord Pos C Power cord safety lock 4 Press the Standby button on the front control panel Fig 8 Front Control Panel Pos A Standby button ...

Page 12: ... on Verify the Standby button changes to green Allow the boot sequence to complete and then verify the correct illumination of the I O module LEDs Fig 9 Successful Boot LED Illumination Pos 1 Off Pos 2 Blinking Pos 3 On Pos 4 On Pos 5 On Pos 6 On Pos 7 On Pos 8 On or off depending on control panel status Pos V On NOTE If the LEDs are not displayed as described contact the Siemens Uptime Service Ce...

Page 13: ...n Fig 10 Blank Screen NOTE The boot sequence takes approximately 80 seconds 150 seconds if POST is set to run during the boot sequence If the system does not boot within 80 to 150 seconds press STANDBY and wait for the system to shut down Verify that the power connections are correct and then boot the system If system does not boot report this to the Siemens Uptime Service Center ...

Page 14: ...on Pos B Keyboard Function keys and Video I O key Pos C Transducer selection L R Flip Pictogram M mode 4B display Text and Dual Display controls Pos D Escape Menu Select 2D Zoom Depth Set Caliper and the Trackball controls Pos E DGC sliders Pos F Focus Split Multi Hertz Cine Digital Store 2 Review Print Store and Digital Store 1 controls Pos G Freeze button See the User Manual for complete descrip...

Page 15: ...ecrease the contrast Turn the Brightness control clockwise for a brighter image with a lighter back ground and counterclockwise to darken Fig 12 Brightness and Contrast Controls 2 Verify the date and time shown in the upper right corner of the screen is accurate Reset the time or date if either are inaccurate Press F6 to display the Preset Main menu Select Day Time from the System Configuration su...

Page 16: ...he Freeze key is back lit in blue when the freeze function is activated CAUTION Freeze the imaging mode prior to connecting or disconnecting a transducer Failure to do so may cause damage to the system or transducer 3 Plug one of the transducers shipped with the system into the upper transducer port If an image does not display press the Transducer button to select the connected trans ducer Positi...

Page 17: ...er cable does not touch the floor Fig 15 Transducer Holder and transducer Cable Holder Pos A Transducer Holder Pos B Cable Holder 4 Press Freeze again to unfreeze the imaging mode 5 Verify that the transducer initializes and that the transducer name is displayed on the screen 6 Repeat the above steps using the lower transducer port Verify the Imaging Modes 0 1 Press Freeze 2 Connect a transducer t...

Page 18: ...Delivery Procedure SONOLINE G20 US06 101 812 01 02 02 Siemens 10 04 sd Page 18 of 26 Medical Solutions USA Inc Fig 16 M mode Screen 5 Press 2D to exit M mode Fig 17 2D Mode Screen 6 Repeat the above steps for each transducer ...

Page 19: ...ted between the printer and the G20 system Verify the AC power cord is connected between the printer and the OEM shelf outlet Fig 18 Printer AC Power Connections Pos A Printer AC inlet Pos B OEM shelf AC outlet 2 Confirm that the printer BNC video and the remote cables are connected between the printer and the I O module Fig 19 Printer BNC Video and Remote Cables Pos A BNC video cable Pos B Remote...

Page 20: ...2 through 6 set to OFF down Fig 20 Printer DIP SW Settings 4 Power on the G20 system and then the printer 5 Press F6 to display the Preset Main menu 6 Select Customize Keys from the Preset Main menu and press the Set key 7 Select B W Print for the Print Store key Fig 21 Customize Keys Submenu 8 Click Save and press Set 9 Press Print Store Confirm that the 2D image displayed on the monitor prints c...

Page 21: ...the image prints NOTE Contact the Siemens Uptime Service Center if there are any prob lems with the printer Verify VCR Operation if installed 0 1 Verify the cable connections between the HS MD3000 S VCR and the G20 system Verify that the AC power cord is connected between the VCR AC inlet and the OEM shelf outlet 2 Verify the VCR cables are connected between the VCR and the I O module Fig 22 VCR C...

Page 22: ...hm S VIDEO switch to ON Fig 23 VCR 75 Ohm S VIDEO Switch 4 Power on the G20 system and then the VCR 5 Press F6 to display the Preset Main menu Fig 24 Peripheral Screen 6 Select Peripheral from the System Configuration sub menu and press the Set key 7 Verify the Video format selections American European Video Format NTSC PAL Video Source S VHS S VHS ...

Page 23: ...tton Pos B Monitor Audio and Key Lock switches Pos C Stop button Pos D Play button Pos E Frame Field Enter button Pos F Jog Adjust dial Pos G Record button Pos H Rewind button 1 Press the VCR Power button to power on the VCR item A 2 Verify the following settings item B 3 Simultaneously press and hold the Stop item C and Play item D buttons for eight seconds this will access the VCR hidden feature...

Page 24: ...exit the hidden menu 6 Press and hold the Menu button item G for three seconds to access Menu 1 101 xx Menu 2 201 xx and Menu 3 301 xx Hidden Feature Setting Picture Control 1 Switching Point SP 45 Position Up Down Ud 0 Position L R Lr 0 Chroma Amplitude CA 3 Hue CP 1 Color Delay CD 3 Brightness Control 0 Contrast Control 0 Black Display AL The Black Display function has three options AL ON and OF...

Page 25: ... The value displays as xx until it is changed To change the values for each menu item rotate the Jog dial Press Frame Field Enter again to save a changed value then rotate Jog to select the next menu item Use the following tables to verify the menu settings Menu Items Menu 1 Values REC TAPE END 101 00 TAPE END 102 00 TAPE IN 103 00 TAPE LENGTH 104 00 NTSH 104 00 PAL VISS 105 01 TIME CODE REC 106 0...

Page 26: ...ously press REC and PLAY The VCR starts recording Con tinue recording for approximately ten seconds On the VCR press Stop The VCR stops recording On the system control panel press 2D The system returns to 2D mode 11 To verify the recorded images play back as they were displayed during imaging Select the Video I O key Press REW to rewind the tape Press PLAY to play back the recorded images 12 Press...