52
HeartView CT
You can activate the “Auto load in 3D” function on the
Examination Card/Auto Tasking and link it to a recon
job. If the postprocessing type is chosen from the pull
down menu, the reconstructed images will be loaded
automatically into the 3D Card on the Navigator with
the corresponding postprocessing type.
On the 3D Card you have the possibility to create for
MPR, MIPthin Range Parallel and Radial protocols which
can be linked to a special series.
For example, if you always do MIP Reconstructions for
a Coronary CTA examination, you load once the images
into the 3D Card. Select the image type (e. g. MIPthin)
and the orientation, and then open the Range Parallel
function. Adapt the range settings (Image thickness,
Distance between the images etc.), hit the link button
and save. From now on, you have a predefined post-
processing protocol, linked to the series description of
a coronary CTA examination.
Exactly the same can be done for VRT presets. In the
main menu, under Type/VRT Definition, you can link
and save VRT presets with a series description.
Some of the Scan protocols are delivered with links
to a postprocessing protocol. If you do not like our
suggestions, please delete the Range Parallel preset or
overwrite them with your own settings.
Summary of Contents for SOMATOM Sensation Cardiac Version A60
Page 1: ...SOMATOM Sensation Cardiac Application Guide Software Version A60 ...
Page 43: ...43 HeartView CT ...
Page 45: ...45 HeartView CT Fig 1 Image reconstruction with a and without b Cardio Sharp kernel a b ...
Page 63: ...63 WholeBodyAngio ...
Page 65: ...65 LungCare ...
Page 98: ...98 Neck Overview AngioCarotid For CTA studies ...
Page 115: ...115 Abdomen ...
Page 123: ...123 Abdomen ...