6BMESSAGES AND DIAGNOSTICS
7.7 Message list of data and command errors (synchronous messages)
Error
no.
Data and operator
errors - Meaning
Description
0
Adjustment command
could not be executed as
distance between
adjustment points is too
small
The adjustment command could not be executed as the
distance between the adjustment points is too small.
The adjustment weight should cause at least a 5% change
of the signal at the load cell output.
1
Instruction could not be
executed due to fault
The instruction could not be executed due to a fault.
The fault must be corrected first.
2
A non-existent or
inactive channel was
addressed
A non-existent or inactive channel was addressed.
The channel may have to be activated.
3
Code not defined
(e.g. cutoff frequency
digital filter, etc.)
The parameter code is not defined (e.g. cutoff frequency
digital filter, etc.).
4
Unknown data record
The data record does not exist.
5
Command not possible
because scale is not
adjusted
The command can not be executed because the scale is
not adjusted. The scale must be adjusted first. The scale is
adjusted when the adjustment digits for the adjustment
zero, the adjustment weight, and the digits for the
adjustment weight are defined/determined.
6
Waiting time of 5 sec. for
adjustment commands
not observed
The waiting time of 5 sec. for adjustment commands was
not observed. This minimum time must be observed for two
consecutive adjustment commands.
7
The adjustment weight is
negative
The adjustment weight is negative and therefore
impermissible.
8 Reserve
Reserve
9 Reserve
Reserve
10 Reserve
Reserve
11 Reserve
Reserve
12 Reserve
Reserve
13 Reserve
Reserve
14 Reserve
Reserve
15 Reserve
Reserve
Table 7-3
List of data and command errors
7.8 Message list of operating messages (asynchronous errors)
If an error generates an operating message, the red SF LED on the front of the
module is illuminated. The incoming and outgoing operating messages are
signaled in the diagnostic alarm.
7-48
SIWAREX U