![Siemens sitrans Instruction Manual Download Page 12](http://html.mh-extra.com/html/siemens/sitrans/sitrans_instruction-manual_374458012.webp)
Page 8
SITRANS AS 100 – INSTRUCTION MANUAL
7ML19985DM02
Accessory Mounting
Temperature Considerations
*maximum range dependent on model.
Notes:
•
If welding, weld must be a continuous bead. Tacking does not provide
sufficient acoustical coupling.
•
Do not arc weld on equipent connected to a SITRANS AS 100. Remove the
SITRANS AS 100 or disconnect electrically to avoid current flowing through
the sensor.
•
If gluing, use Loctite 326 adhesive, or equivalent. Follow manufacturer’s
instructions to ensure proper adhesion.
Warning: Temperature at Sensor must not exceed minimum or
maximum ratings.
Extension Tab
Screw Sensor into threaded hole of tab,
and fasten onto device being monitored.
Mounting Disc
Screw Sensor into disc, after welding or
bonding disc onto device being monitored.
Note:
for standard Temperature
range model, maximum
temperature is 80
°
C.
Note: If the flange temperature is 100
°
C
and the ambient temperature is 20
°
C,
the Sensor temperature at the
electronics is below the maximum
rating.
Ensure adequate isolation from hot
surfaces by providing additional
spacing.
Tab acts as heat sink.
65 ºC
100 ºC
< 80 ºC
>80 ºC
20 ºC