Commissioning
6.1 Operating via the handheld programmer
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
39
6.1.3
Programming
Note
•
While the device is in PROGRAM mode the output remains active and continues to
respond to changes in the device.
•
The device automatically returns to Measurement mode after a period of inactivity in
PROGRAM mode (between 15 seconds and 2 minutes, depending on the menu level).
Change parameter settings and set operating conditions to suit your specific application. For
remote operation see Operating via SIMATIC PDM (Page 59).
Note
For the complete list of parameters with instructions, see Parameter Reference (Page 93).
Parameters are identified by name and organized into function groups. See LCD menu structure
(Page 249).
1. QUICK START
2. SETUP
2.1. IDENTIFICATION
............
2.6. LINEARIZATION
2.6.1. VOLUME
2.6.1.1. VESSEL SHAPE