SITRANS LH100 (7MF1572..) Transmitter for hydrostatic level
30
A5E32588610-02, 02/2014
4.3
Setting up the measuring points
①
Cable box
③
Pressure transmitter
②
Anchor clamp
Figure 4-2
Basic procedure for setting up the measuring points
4.4
Establishing the measuring range
Calculating the measuring range with process media with a density
≠
1000 kg/m3 (process medium
≠
water)
p =
ρ
* g * h
with:
ρ
= density of the process medium
g = local gravitational acceleration
h = maximum level