2
43-TV-25-20 GLO Issue 4 01/02 UK
A5E00117351-04
Cables
To fully comply with the requirements of the CE Mark, all cables connected to the rear
of the unit should use screened cable terminated at both ends. Also a low impedance
earth cable (<1 m
Ω
) should be connected to the earthing stud on the rear of the recorder.
WARNINGS AND SAFETY PRECAUTIONS
Do’s and Don’ts
1.
Before any other connections are made to the recorder, the protective earth terminal
should be connected to a protective conductor.
2.
In order to comply with the requirements of safety standard EN61010, the recorder
should have one of the following as a disconnecting device, located within easy reach of
the operator, and be clearly labelled as the disconnecting safety device:
• A switch or circuit breaker which complies with the requirements of IEC947-1 and
IEC947-3.
• A separable coupler which can be disconnected without the use of a tool.
• A separable plug, without a locking device, to mate with a socket outlet in
the building.
3.
Whenever it is likely that protection has been impaired, the recorder should be made
inoperative and secured against operation. The manufacturer's service centre should be
contacted.
4.
Any adjustment, maintenance, and repair of the opened recorder in a powered condition
is hazardous and should never be attempted.
5.
Where conductive pollution such as condensation or conductive dust is present, ade-
quate air conditioning, filtering and/or sealing must be installed.
6.
This recorder contains a battery which must be treated and disposed of with care. Bat-
teries must not be short circuited. Batteries should be disposed of in accordance with
local regulations, they must not be disposed of with normal refuse.
See “Appendix B -
Battery Safety Data Sheet” on page 85.
IMPROPER INTERRUPTION OF CONNECTIONS
Any interruption of the protective conductor outside the recorder, or disconnection of
the protective earth terminal is likely to make the recorder dangerous under some
fault conditions. Intentional interruption of the protective conductor is dangerous.
Failure to comply with these instructions will result in death or serious injury.
DANGER
Summary of Contents for SIREC DS 7ND4120 Series
Page 8: ...viii 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 12: ...4 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 22: ...14 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 34: ...26 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 Power Up Screen SIREC DS 0 1 ...
Page 47: ...43 TV 25 20 GLO Issue 4 01 02 UK 39 A5E00117351 04 2 4 4D4 4D4 4D4 E A F G H I 4 49 ...
Page 48: ...40 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 5 SIREC DS 5 4 2 SIREC D Server 4 4 1 4 0 ...
Page 58: ...50 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 0 B 1 4 4 9 9 Digital filter 0 9 9 9 9 ...
Page 66: ...58 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 0 3 7 3 0 3 0 0 0 0 A 0 0 4 5 0 G 0 ...
Page 70: ...62 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 76: ...68 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 2 2 2 4 2 Validate Disk ...
Page 78: ...70 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 80: ...72 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 86: ...78 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 90: ...82 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 92: ...84 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 95: ...43 TV 25 20 GLO Issue 4 01 02 UK 87 A5E00117351 04 4 45 4 5 C B 4 0 A 4 4 H F 4 HG G 4 ...
Page 96: ...88 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 100: ...92 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 104: ...96 43 TV 25 20 GLO Issue 4 01 02 UK A5E00117351 04 ...
Page 115: ...43 TV 25 20 GLO Issue 4 01 02 UK ...