Connect the gas lines to the corresponding inlets/outlets.
WARNING
Introduction of toxic, aggressive or flammable gases
When toxic, aggressive or flammable gases are introduced, carry out a leak test for these
pipe connections.
You can find additional information in the section →Checking the sample gas path for leaks
(Page 74).
Note
Pipes
Connect the pipes to the gas connections. Use appropriate clamping rings with union nuts for
a pipe diameter of 6 mm.
The material of the parts in contact with the gas (threaded joints and clamping rings) must be
appropriate to the application (stainless steel or Hastelloy).
6.3.2
Connecting the OXYMAT 7 field module
Overview
Figure 6-2
Field module schematic
Figure 6-3
End of field module cable (schematic)
Connect the end of the field module cable to the terminal block of the field control unit. The
preassembled cable on the field module is fanned out in the delivered condition and provided
with end sleeves. The cable length is 7 meters.
The core assignments and the associated connection numbers of the terminal block are shown
in the following table. You can find the terminal block layout in the section → Field control unit
Wiring
6.3 Field module
Field device
52
Operating Instructions, 05/2015, A5E35640463-02
Summary of Contents for SIPROCESS GA700
Page 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Page 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Page 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Page 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Page 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...