6-68
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D41 AQC Cartridge Prime Error: 1, 2, 3
D42 R Cartridge Prime Error: 1, 2
Problem
Corrective Action
QC material failed
to flow from a new
AutomaticQC
cartridge.
1 level 1
2 level 2
3 level 3
1. Check that the AutomaticQC connector is completely
closed.
2. Prime the AutomaticQC cartridge.
3. Perform a leak test.
4. Check if there is an obstruction in the reagent probe
cartridge.
Refer to
Flushing the Tubing for the AutomaticQC Module
5. If the problem persists, replace the AutomaticQC cartridge.
NOTE:
If you are not able to replace the cartridge, in Setup,
turn the AutomaticQC option off to continue analyzing patient
samples.
6. If the problem persists, call for technical assistance.
Problem
Corrective Action
200 cal or slope cal
reagent failed to
flow from a new
reagent cartridge.
1 200 cal reagent
2 slope cal reagent
•
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
•
Remove and reinstall the reagent cartridge.
Refer to
Ejecting the Wash or Reagent Cartridge
•
Perform a leak test.
Refer to
‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
•
If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
•
If the problem persists, perform a 2-point calibration.
•
If the problem persists, call for technical assistance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...