6-20
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
b. If using a standard syringe, hold the syringe firmly.
5. Press the syringe plunger gently while holding gauze against the port at the back of the
sample connector (item 1, Figure 6-7), to catch the obstruction and any excess liquid.
6. Remove the syringe.
7. Discard the gauze.
Figure 6-7 Reagent Manifold and Tubing Connections
8. Reinstall the reagent manifold.
a. Insert the reagent manifold until it latches into place, using the 2 latches (item 5,
Figure 6-7) that secure the manifold to the system.
CAUTION:
If the path is totally obstructed, liquid may be expelled suddenly and
with great force through the port at the back of the sample connector.
BIOHAZARD:
Discard the gauze according to the established biohazard disposal
practices of your workplace.
1
Fitting to which the reagent manifold tubing and CO-ox pump tubing are connected
(Rapidlab 1245 and 1265 systems)
2
Reagent manifold tubing that connects the fitting to the port on the top, left-side of the
reagent manifold (Rapidlab 1245 and 1265 systems)
3
Sample connector
4
Waste tubing connected to port on top, right-side of the reagent manifold
5
Two latches that secure the reagent manifold to the system
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...