32
Safety Inspection
POWERMOBIL
SPR2-240.831.01.02.02
Siemens AG
07.05
CS PS 24
Page 32 of 44
Medical Solutions
NOTE
•
Depending on the programming, the audible warning signal
has to sound 4.5, 5 or 9.5 minutes after the fluoroscopy has
started. It sounds again every 5 or 10 minutes. It is turned off
by pressing the -0- button once.
•
Pressing this button again resets the fluoroscopic timer to "0".
•
Depending on the country-specific programming, either the re-
quired radiation shut off does not occur or it occurs every 5 or
10 minutes and then every 5 or 10 minutes after that.
SIE
Iris collimator
•
Check the iris collimator and correct it if necessary.
X
- Select the full I.I. format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIE
Area-dose product measuring device
•
Check the area-dose product measuring device
X
•
(see the POWERMOBIL service instructions for this purpose).
SIE
Codonics printer cable breaks
•
Inspect all connection lines between the printer and the Framegrabber for damage and
breaks. Inspect the cable routing of the energy chain for damage and proper functional-
ity (ensure that none of the cables are pinched).
3.2.1 Monitors/displays
SIE
Monitor/displays
•
Check the SIMOMED N monitor as prescribed in the checklist.
•
Check the luminance of the standard monitors with 100/120 Hz.
•
Check the luminance of the TFT displays. Check the ambient light sensor.