28
Safety Inspection
POWERMOBIL
SPR2-240.831.01.02.02
Siemens AG
07.05
CS PS 24
Page 28 of 44
Medical Solutions
3-
3
Safety Inspection
3.1
Mechanical safety
SIM
Cover panels
•
Remove the covers from the POWERMOBIL stand and monitor trolley.
•
Check cover panels for damage.
SIM
Cable deflectors
•
Inspect the cable deflectors on the stand and on the monitor trolley and replace as
needed.
SIM
Foot brake
•
Test the function of the foot brakes on the POWERMOBIL stand and monitor trolley on
a level floor.
SIM
Brakes
•
Use the spring balance to test whether the orbital, angulation, swivel and horizontal
movement brakes apply the prescribed force. (Refer to service instructions POWER-
MOBIL).
SIM
C-arm
•
Check the C-arm for damage to the carbon fiber structures (visual inspection). If there is
damage, proceed as described in the warning in chapter 1.
•
Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
•
Orbital movement: Once the orbital brake is released, the C-arm can move a maximum
of 5° on its own. Check the servo support function of the orbital movement. Check
whether the orbital movement is smooth.
SIM
Wheels
•
Move the POWERMOBIL in a straight line on a level floor.
•
Evaluate the straight and quiet movement of the POWERMOBIL.
•
Replace any defective wheels or castors.
•
Check the steering function.
SIM
Lifting column
•
Switch the system on
•
Electrically raise and lower the lifting column,
- while listening for noises and checking for play in the bearings.
- The lifting column movement must switch off automatically when the end positions
are reached.
•
Switch off the system.
•
Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIM
Emergency Stop switch
•
Press the Emergency Stop switch. The lifting column may not be raised or lowered.