Page EN-14
Pointek ULS200 – QUICK START MANUAL
A5E32268616
mmmmm
English
Output Function
The alarms can be set to function in the desired mode.
* Factory setting
Function display:
Setpoints
Function
Relay 1
Relay 2
1 *
high alarm
low alarm
2
high alarm
high alarm
3
low alarm
low alarm
4
high alarm
LOE alarm
5
low alarm
LOE alarm
6
pump down
low alarm
7
pump down
high alarm
8
pump up
low alarm
9
pump up
high alarm
10
pump down
LOE alarm
11
pump up
LOE alarm
12
pump down
pump up
13
pump down
pump down
14
pump up
pump up
high alarm
H
high-high alarm
HH
low alarm
L
low-low alarm
LL
loss of echo alarm
LOE
pump up control
PU
pump down control PD
The setpoints can be set where reference levels,
either from the material in the vessel or from a
target, cannot be provided. This method can also
be used to trim the output levels obtained by the
Reference Method (Quick Start).
The setpoints are referenced from the face of the
sensor. They should not be set at or above the
blanking value, or at or below the range limit.
Factory Setting: Relay 1 = 0.5 m (1.64 ft)
Relay 2 = 4.50 m (14.76 ft)
high-high
high
low
low-low
ULS200
setpoint
blanking
range
limit