5 - 1
Siemens AG
RX57-029.061.01
Page 1 of 2
MEMOSKOP 2/3/50/100/SUB
Medical Engineering
Rev. 02
10.97
TD SD 34
Electrical settings/programming
5
SIREMOBIL 2000
5
•
For the configuration of the MEMOSKOP 3/50/(100)
refer to chapter 5 of the service instructions RR2-120.061.02.. .
•
For the configuration of the MEMOSKOP SUB 2000
refer to chapter 5 of the service instructions RR2-120.061.02.. .
UROSKOP D1/D2
5
•
For the configuration of the Memoskop 50 / (100)
•
Switch the system "ON"
•
Wait for the MEMOSKOP to boot.
Activate set-up mode
5
•
Actuate the following keys on the keyboard in sequence:
- <NEXT PAT>
- <@>
- <%>
- <CTRL> + <S>
•
After activating set-up mode, a sequence of numbers is displayed in the lower region of
the monitor image.
•
Enter the values indicated for the relevant TV system on the keyboard:
- for TV systems with 60Hz and 120 Hz memory output:
02 00 66 08 FE FF FF 01 02 32 63 63
- for TV systems with 50 Hz and 100 Hz memory output:
02 00 66 08 FF FF FF 01 02 32 63 63
•
Enter the numbers without any blanks.
•
Press the keys <CTRL> + <E>.
Exit set-up mode
5
•
Switch the system "OFF" and "ON" again.
The memory returns to normal operating mode.
If PC board D18 on MEMOSKOP 2K or PC board D1 or D2 on
MEMOSKOP 50/(100) is replaced, a download of the MEMOSKOP
software must be performed on UROSKOP.
If PC board D1 is being replaced in MEMOSKOP 50/(100), the
MEMOSKOP must be configured additionally via the keyboard.
<x> means: press key x.
<CTRL> + <E> means: press and hold <CTRL> and simulta-
neously press key <E>.
CAUTION
NOTE