Checking dose settings
8 - 7
Siemens AG
SPB7-250.815.02
Page 7 of 10
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Check the H & D dose settings for Mo/Rh and W/Rh.
8
If the Mo/Mo AEC does values for
H & D
are within the specification, proceed as follows:
1.
Mount the Collimator Mounted Plexi Phantom.
2.
Select the appropriate exposure parameters according to the table and start an
exposure.
3.
Select the anode/filter combination according to the table and set the mAs (control
console) as close as possible to the H & D mAs values (Mo/Mo) that you have
measured in step 2.
Display the image with the
syngo Image Viewer
. To do so click on the
Viewing
tabcard, load the image with the
Patient -> Browser
and select
Tools -> Rect-
angle
.
Draw an area of approximately
1 sq.cm
and measure the pixel mean
value.
Add the result to the table.
The Mo/Rh and W/Rh pixel values must be as close as possible to the Mo/Mo
pixel values (see table at step 2).
The deviation must not /- 20 %.
4.
If the
H & D
dose settings have to be adjusted, proceed with Chapter 8 "Adjust the
H and D Settings" and repeat the
H & D
dose checks for Mo/Rh and W/Rh again.
5.
Record the measured values in the
Anode /
Filter
AEC
kV
Pixel value
Mo/Rh
H
28 kV
D
28 kV
W/Rh
H
28 kV
D
28 kV
Tab. 2
Checking the H & D dose settings for Mo/Rh and W/Rh
"