
Mechanical Checks
5 - 9
Siemens AG
SPB7-250.815.02
Page 9 of 12
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Adjusting the field light time
5
The illumination time of the light field can be adjusted within a certain range according to
the customer´s requirements.
1.
Connect the service PC to the D750 and start the MAMMOMAT Service Software.
2.
In
Mainmenu
select
Configuration
Miscellaneous
Illumination time
.
3.
Type in the illumination time and save with
<F2>
.
Checking the compression device
5
1.
Attach the compression plate (1/Fig. 10) if you have not already done so.
2.
Check that the compression plate cannot be removed without pressing the
compression plate release button (2/Fig. 10).
3.
Check the function of the knobs for manual compression/decompression
(3/Fig. 10).
4.
Place a piece of soft material on the object table, for example a rolled up towel or a
piece of foam rubber (4/Fig. 10).
A shorter illumination time results in an increased life time of the
lamp.
Fig. 10 Swivel arm system
NOTE
1
2
3
5
4