
MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 8 of 12
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
5 - 8
Mechanical Checks
Checking the field light time
5
1.
Switch on the field light by pressing the switch (1/Fig. 8) or by pressing the
compression foot switch (1/Fig. 9).
The field light shall go out automatically after a preset period of time (normally
20 s
.).
Fig. 8
Field lightswitch
Fig. 9
Foot switch
1
MA
M00
133
1