Detector installation
Adjusting the switches
4
1.
Mount the micro switches S2 and S3 (Fig. 8) along the markings.
2.
Adjustments of S2
Move the grid to the front. S2 must switch 2 mm (+1 mm tolerance) before the grid
reaches the end position.
3.
Adjustment of S3
Move the grid to the back and insert the magnification table. S3 must switch 2 mm
(+1 mm tolerance) before the magnification table reaches the end position.
Mounting the covers
4
1.
Mount the bottom wing cover with 6 screws.
2.
Mount the top wing cover with 4 screws.
3.
Slide the detector fibre cover carefully over the detector and fasten it with 6 screws
at the bottom of the cover.
Fig. 8
Micro switch installation
The magnification table is optional. When no magnification is
supplied with the system, the micro switch has to be mounted
along the markings.
Remember to check the switch adjustment if the customer orders
a magnification table.