Mammomat 1000/3000
Register 03
SP B7-230.033.07
Page 4 of 8
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
3 - 4
Preparatory work
2.
Cut and remove the cable tie securing the balancing spring during transport.
Transport safeguard for the rotary motion
3
1.
Remove the transport safeguard screw (1/Fig. 4) with spacer to enable rotary
motion.
The screw and the spacer should be kept for the mounting of the lifting carriage
transport safeguard.
2.
Mount the red lifting carriage transport safeguard with screw and spacer
(1/Fig. 4) on the left hand side of the stand, see Fig. 5. Insert the screw through the
slot from the inside of the stand, see Fig. 6
3.
Tighten the screw. Make sure that the transport safeguard is parallell to the
stand.
The lifting carriage transport safeguard shall be kept for service
purposes (Fig. 5).
Fig. 4
Transport safeguard for the rotary motion
Fig. 5
Transport safeguard mounted on the stand
Fig. 6
Screw and spacer inserted from the inside of the
stand
NOTICE
1
MA
M0
0083
MAM00312
MAM00313
MAM00314