LaboStar 7 TWF DI and UV
Ultra pure water system
Catalog no. 2231 & 2232
Page 17 of 26
4.8
Monitor
监
监
监
监控
控
控
控
During operation, the monitor displays the current system status.
Conductivity value Q2 for product water is shown in the left part of the monitor. The right part displays either
conductivity value Q1, temperature or the level in the tank.
在正常运行状态下,纯水机的监控屏幕显示状态如下图:
监控屏幕左侧
显示
Q2
,它代表产水水质的电导率;
监控屏幕的右侧显示
Q1(
反渗透掺水
)
、产水温度、水箱液位。
••••
Press the arrow keys [
] to switch between temperature, Q1 and tank level in the right part of the
display.
通过上
/
下箭头选取想要的显示参数
- Q1
、产水温度、水箱液位。
Tank display:
水
水
水
水箱
箱
箱
箱显
显
显
显示
示
示
示
Tank empty:
Active protection to prevent the pump from running dry; water extraction is
not possible.
< 50 %
Circulation pump starts up, product water extraction is now possible.
> 50 %
Once the tank is > 50 % full, RO production continues for a few minutes (see
Chapter 5.3 - Tank settings). Once the cycle is complete, production stops
and the RO system switches to stand-by mode. The Q1 measurement
displayed in the monitor is on standby (Stdby) until enough water is drawn to
lower the tank level to < 50 %. At this point, RO production restarts.
空水箱:
不能向外供水,防止水泵无水运行
-
干转;
< 50 %
循环水泵开始运行,
可以向外供水;
> 50 %
一旦水箱液位
> 50 %
,
RO
反渗透还会继续产水几分钟(参见
5.3
章节
-Tank
setting
)
当这个循环周期结束后,
RO
反渗透停止产水转向备用状态,监控屏幕右
侧不再显示
Q1
值,而是显示“
Stdby
”即备用状态,一直到水箱内的液位降至
<
50 %
,
RO
反渗透又开始运行。
Current
measured value
Temperature, conductivity Q1 or
tank level display
Units
Product water