Guidance and manufacturer's declaration -
electromagnetic immunity
The HearCheck Screener is intended for use in the electro magnetic environment
specified below. The customer or the user of the HearCheck Screener should
assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV
differential
mode
± 2 kV
common
mode
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.