Functions
8.3 Calibration
FIDAMAT 6
82
Operating Instructions, 01/2019, A5E00222135-04
With a total range calibration, you calibrate all measuring ranges together. Specify the
"leading" measuring range under function 22. We recommend using the largest measuring
range for this.
The setpoint and current actual value of the "leading" measuring range appear in the display.
When the actual value has stabilized, you can initiate the calibration by pressing the fourth
softkey. The actual value is now made to agree with the setpoint.
If you make a mistake during the calibration (e.g. wrong calibration gas), reload the original
calibration by pressing the "CANCEL" softkey.
See also
Setting the setpoints (function 22) (Page 82)
Total / single calibration (function 23) (Page 83)
Define measuring ranges (function 41) (Page 93)
8.3.4
Setting the setpoints (function 22)
Figure 8-6
Setting the setpoints
The function screen shows the setpoint input with total calibration. The fourth measuring
range is chosen here as the leading measuring range.
With single calibration, there is no choice for the leading measuring range.
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......