Table 11-8
Ex conditions: Max. temperature vs. temperature classifaction (T amb: -20°C to +50°C)
DI 1/1.5/2.1/3
DI 6/15
T6
75
49
T5
100
100
T4
135
135
T3
180
180
See also
Pressure - temperature ratings (Page 122)
11.3
Construction
Table 11-9
Designated use
Description
Specification
Measurement of process media
• Fluid Group 1 (suitable for dangerous fluids)
• Aggregate state: Paste/light slurry, liquid and gas
Table 11-10 Function and system design
Description
Specification
Measuring principle
Coriolis
System architecture
• Point-to-point (1 Modbus RTU master - 1 trans‐
mitter slave)
• Multidrop (1 Modbus RTU master - up to 32
transmitter slaves per segment. Max 247 slaves)
The MASS2100 and FC300 7ME4817-... is preconfigured for integration with the SITRANS
FCT070.
Note
NPT glands
When using NPT glands, user must take care when packing threads and installing cables that
sufficient tightness is obtained to prevent ingress of moisture.
Technical data
11.3 Construction
SITRANS MASS 2100 & FC300 (FCT010/DSL)
110