6.1.1
Warnings
CAUTION
Sensor and transmitter ordered separately
If the sensor and the transmitter are ordered separately, a "Set To Default" routine must be
performed. This can be done via SIMATIC PDM or via menu item 3.3.3 in HMI.
WARNING
Dangerous high voltage
Certain parts inside the device carry dangerous high voltage. The housing must be closed
and grounded before switching on the device.
WARNING
Improper handling
The sensor connected to this device can be operated with high pressure and corrosive media.
Therefore improper handling of the device can lead to serious injuries and/or considerable
material damage.
6.2
General requirements
Before commissioning it must be checked that:
● the device has been installed and connected in accordance with the guidelines provided in
Installing/Mounting (Page 37) and Connecting (Page 57)
● device installed in hazardous area meets the requirements described in Use in hazardous
If the sensor and the transmitter are ordered separately, a "Set To Default" routine must be
performed. This can be done via SIMATIC PDM or via menu item 3.3.3 in HMI.
6.3
Local commissioning
6.3.1
Power-up
Power up the device. Devices with local display show a screen for initial startup.
6.3.2
Local display
The device is commissioned/operated with the touch keypad on the local display.
Commissioning
6.3 Local commissioning
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
93