CLINITEK Analyzer Operator’s Guide
33
Op
erations
Viewing Sample Interference Notes
Sample interference notes inform you about the test results that can
be affected by components detected in the urine sample. By default,
sample interference notes display and print. To set up sample
interference notes, see Section 7,
Depending on the strip and sample, sample interference notes could
include the following statements:
•
High SG may cause falsely lowered GLU results.
•
Elevated GLU may cause falsely lowered LEU results.
•
Visibly bloody urine may cause falsely elevated PRO results.
•
High SG may cause falsely lowered LEU results.
•
High pH may cause falsely elevated PRO results.
Printing the Urinalysis Strip Quick Test Results
Print the urinalysis strip Quick Test results manually or automatically,
or send the results to a computer.
To print the urinalysis strip Quick Test results manually, perform the
following steps:
Select
to print the test results.
•
The date, time, test sequence number, and test results display
on the printout.
•
For Color and Clarity, the value is
Not Entered
.
•
If the results are positive, an asterisk (
*
) displays next to the
results, only if you selected
Mark positive results
in
Instrument Set Up.