CLINITEK Analyzer Operator’s Guide
119
System C
onf
iguration
2. Select
Yes
.
Note
If you want to keep the operators list, select
No
. The
Authorized operator
screen displays.
The system deletes all the operators in the list and disables the
Authorized Operator setting.
Setting up Sample Interference Notes
Sample interference notes inform you about test results that can be
affected by components detected in the urine sample. By default,
sample interference notes display and print.
Depending on the strip and sample, sample interference notes could
include the following statements:
•
High SG may cause falsely lowered GLU results.
•
Elevated GLU may cause falsely lowered LEU results.
•
Visibly bloody urine may cause falsely elevated PRO results.
•
High SG may cause falsely lowered LEU results.
•
High pH may cause falsely elevated PRO results.
To set up sample interference notes, perform the following steps:
1. On the
Select Ready
screen, select
Instrument Set Up
.
2. On the
Choose Settings
screen, select
Instrument Settings
.
3. On the
Instrument Settings
screen, select
Sample Interference
Notes
.
4. On the
Notes Settings
screen, perform the following steps:
•
To include sample interference notes, select
Enabled
(default).
•
To exclude sample interference notes, select
Disabled
.
5. Select
Done
twice.
Changing the Printer Settings
You can customize the printed test results by including or excluding
the following printout options. By default, all printout options are
enabled except the
Custom Header
option, which is disabled. You also
need to configure the printout options.
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...